The Uni-T UT181A True RMS Datalogging Multimeter is…a terrible data logger (WHL #124)
Everybody runs searches on eBay regarding nice-to-have equipment, right? So do I, and in terms of datalogging multimeters, one of them showed a hit last week. A device that is actually rarely offered, and at a reasonable price as well – a Uni-T branded UT181A “True RMS Datalogging Multimeter”. So I got it. And, well, in case you missed the headline: It’s a pretty decent meter, but it fails at the one thing that it was specifically designed to do: Logging data.
But first things first, this is the device in question, with a couple of what we call Wurstfinger, sausage fingers, reflecting on the display surface. A side effect of switching to the much bulkier Sony Alpha 6100 compared to my old (and now sold) RX100M2.
This is a chunky boi, and it is exactly the same case size (with different buttons, so different front moulding) as my Uni-T UT505A 1200V insulation meter. That one is 658g, my also quite bulky Greenlee DM-860A = Brymen BM869s is 649g, and this UT181A is 602g. For comparison, the Aneng AN8008 pocket meter is only 136g, so not even 1/4 the mass.
It does come with a built-in battery which is handy, and it cannot be charged during use, which is extremely un-handy and a total fail of the product designers as well. It’s also classified as IP65, but I’m not going to test that…the back shell water-proofing relies on the 3×2 screws at the red-black moulding corners, plus three large captive and two significantly smaller screws in between that hold the battery compartment in place. They probably found out that the flex of that span of plastic is a bit too much to achieve the IP rating and added the smaller ones later on.
The battery compartment (the smaller screws might compromise the seal a couple years down the line, check how they cut into it). One can gain access to the two HRC fuses here (good), but not the CR2032 battery that is likely used for the clock (bad!), more on that later.
Further disassembly:
Main ICs:
* ST STM32F103ZEt6 (the brains™)
* ISSI ISSIWV1024168LL-10TLI (16Mbit = 2MB SRAM – not sure if they buffer readings via the battery?)
* ST NAND512W3A2SN6 (512Mbit = 64MB NAND flash, most likely for firmware and not for user data)
* RAiO RA8875L3N 1630-N STOS02 (the display controller)
* Cyrustek ES51997P 1440-GZ7X (DMM analog frontend, according to the interwebs also used in the Voltcraft VC890 which coincidentally also was, but no longer is on my eBay shopping list)
Generally nice construction with decent circuit protection, but I’m not doing the Dave Jones here and trace all potential HV paths and measure their creepage distance. I’ve seen better, but I have seen much more worse examples than this.
And even further disassembly: The front name plate is just a single insertable frame and the button connection is done via a flat flex that is semi-difficult to reinsert. Very satisfying rotary clicking mechanism.
Front side of the main board: There are quite a few components, but mainly passives and small aux active components, no major ICs. The display connection is done via a populated flat flex that is awkwardly folded (and then held in by the name plate frame, which needs to be mounted into the PCB cutouts and then tightened first, trap for young players!)
Internal construction almost aside: Charging. My used unit did not come with the original AC plug and charging adapter, so I can only show the press material – that’s the large dark grey thingy with the center hole (the one in front of it is the adapter for the two thermocouples) in the corner of the bag and the cardboard packaging.
Charging without that is easy – apply 9.5V at 500mA max to the COM (minus) and mA (plus) port once the device is set to charging mode. 9V is possible, although it gets very slow at the end and does false trigger 100% state of charge towards the end while still charging. The circuitry does seem to check current and declares “full” once it drops below like 100mA, which apparently was an issue for early models with a 9.0V instead of 9.5V (or 10.0V?) wall adapter. The internal battery is a regular 7.4V (so 2S) prismatic lithium cell, which tops out at 8.4V and therefore the charging LDO only has about 0.6V of headroom, which doesn’t seem enough. Do not apply significantly more voltage, e.g. similar Fluke meters charge on a 24V supply…
My issue with charging is that although the meter has a great battery life, I expect a logging meter to run 24/7 if needed. And that is not possible with that charging setup, as the regular charging adapter covers all ports regardless (safety, bla, bla), and even if injected just into the two relevant banana jacks, it has to be in a charging mode that is mutually exclusive with logging. So once the battery is depleted – game over. Uni-T claims 2000mAh of battery capacity but fails to convert that into run time hours in both the datasheet as well as the manual. Anyway, the display dims and then shuts off after a user-defined time, saving precious electrons, so it can do hours of logging, but certainly not days and weeks. Some guy on the EEVblog forums tried a Qi charging pad, as the case does offer a little unused space inside. Might work and is an elegant way to keep the case water tight, but a simple barrel jack would have been preferred.
Next logging disaster (actually a double whammy): Uni-T advertises an iOS/Android logging app.
Fuckup #1: The required bluetooth adapter is an optional extra.
Fuckup #2: The app is plastered with advertising that leads to HTTP404 errors on the Uni-T website, and requires an account before it wants to connect to any device.
Horse shit. Unmaintained trash that just exists because it was decided in a board meeting, not because they really want to support that kind of measurement (or they would maintain their app and publish the connection details, API calls and whatnot once they put it to EOL status, right?)
Sure, there’s an optical USB link as well, but I don’t have that module either (included as standard, though), and it likely only does work on Windows. I do have some trust in the community that somebody had a look at this and made an open source version, but since I won’t keep the meter, I didn’t bother.
And now my main gripe with the “logging” part: Yes, the meter does log, up to 20k points in user-defined time steps. It also does show several things during logging like min, max and average value – glad someone took advantage of the big 320×240 graphical display. But…it cannot show a graph of the acquired data DURING the measurement. It can do that afterwards, but then it cannot resume the measurement. You have to stop logging to see the logged data. Why. The. Fuck?! Who thought this is a great idea? Who signed up on this and said “ship it!”?
And even that graphical view is, well, limited, if not bugged. One can scroll horizontally, so view chunks of data of longer measurements. Great. But it is not possible to scroll vertically, one can only scale that axis. So when doing like AC mains voltage logging, the only option is to have an empty graph that shows say 0-10VAC, or to have a pixel-perfect straight line when fully zoomed out to see 0-600VAC. That, again, is utter horse shit. Nobody thought of this, nobody cared, and this is still sold as a “datalogging multimeter”. Why do we have a power hungry 3.5″ color display if we don’t use it?! At currently 488 USD MSRP, this is supposed to be a standalone device, and not require a PC or phone for viewing the acquired data all the time!
Final bits: My (again, used) unit has a column of defective pixels, which is why, in addition to all of the logging disappointment, this unit will go back to the commercial seller (it was technically shown in a blurry image, but not described; instead, it was sold as “in good condition, display even has the original protective film applied”). If it were logging as it should, I would have kept it with that defect, it doesn’t really bother me at the edge…but it’s not a logging meter, just one more regular meter that will gather dust because I already have plenty of regular meters.
It also has a perfectly working clock when powered on, which will basically fully stop working the second the device is shut off. This should be fixable with a new CR2032 cell…but why does it have a separate battery in the first place? And why doesn’t it lose the entire date and time information if it is fully depleted? Why is the CR2032 not accessible without fully taking the unit apart, but the not-to-be-user-replaced main battery is? What another load of strange design decisions.
I also did test this meter for use as a meter – mind-boggling, I know.
At my voltage test unit, these were the results:
2.5V range – reference reading on a 6 1/2 digit meter was 2.49887V, UT181 says 2.4988V (only a 60000 count meter!), that’s -0,003% if calculating with the full accuracy and forgetting about the missing digit
5.0V range – ref 5.00162V, UT181A 5.0018V (+0,004%)
7.5V range – ref 7.49990V, UT181A 7.499V (-0,012%)
10V range – ref 10.00217V, UT181A 10.000V (-0,022%)
Datasheet accuracy on DC voltage 0.025% plus 5 digits, so perfect all along.
Resistance (via the 7 decade thumbwheel resistor box (project #P1)), testing with digit “5” to maximize the counts, as well as to omit the 1R resistor that deviates more than 1% from the intended value due to contact resistance and whatnot:
0R cable-cable: 0.00R spot-on (using the average feature for all measurements)
0R including box at 0R: 2.46R (moving, not nulled out – the meter can display the measured and the corrected value simultaneously until switching ranges, great feature!)
1R: 2.47R (I think my 0R + 1R link is bad!)
5R: 6.30R (so 3.83R increase on adding 4R, that sounds like a box problem)
55R: 55.05R (+0,29 %)
555R: 554.26R (-0,23%)
5k555: 5k5519 (+0.25%)
55k555: 55k70 (+0,06%) (why is it only four digits? it switches at 55k4 into the next range! but it doesn’t do this on the others? W-T-F?)
555k555: 554k70 (-0,05%)
5M555555: 5M5524 (-0,06%)
Resistance accuracy is advertised as 0.05% plus 2 digits. I’d say that is generally met and my box is just out – it does meet specs nicely on single ranges and it is out like fivefold on others. I haven’t tested the box in ages, maybe I should do that again)
Short circuit beeper test is basically instant, I would not complain about that (picky as Dave and testing the same way)
And last, but not least: Diode compliance (test) voltage is a handy 3.14V, and 3.1eV equal 400nm in wavelength. That’s not exceptionally high given the 7.4V nominal / 8.4V max battery, but it should be good for any single-chip LED up to literally the beginning of the UV spectrum.
Final verdict: Great meter, shitty logging meter. Since it is explicitly marketed as a logging meter and it only has a graphical display because of this: It is unfortunately a pretty shitty one, 2/5, wouldn’t recommend…
(If you only need a meter and logging is a bonus: That’s a 4/5 rating, if you can stay well below MSRP 😉)